Laboratory separator-train mass audit
Use pvt_separator_mass.audit_separator_mass for laboratory tests where oil
passes through a sequence of lower-pressure separators, stage gas is removed,
and the remaining oil becomes the feed for the next stage. Use the
DLE audit for a differential liberation pressure series.
This source check does not run a flash or fit fluid parameters.
Input and basis
Use protocol SEPARATOR and supply source references for the sample,
experiment, stock-tank oil, data-use terms and interpretation of the bounds.
The first row describes the incoming oil. Subsequent rows describe the oil
after each stage and the gas removed at that stage. The last row must be the
actual stock-tank oil reference used to normalize all volumes. Each row needs
an ID, a source reference and its own conditions. Absolute pressure intervals
must decrease without overlap. Temperature can vary between stages.
| Source field | Input unit |
|---|---|
pressure, temperature |
bara, K |
oil_volume_ratio |
m3/m3 stock tank oil |
oil_density |
kg/m3 at the row conditions |
solution_gor, released_gas |
Sm3/m3 stock tank oil |
standard_gas_density |
kg/Sm3 for the stage’s released gas |
released_gas_molar_mass, standard_gas_z |
kg/mol, 1 for the released gas |
standard_gas.pressure, standard_gas.temperature |
Pa absolute, K |
Supply each quantity as a closed interval with lower and upper decimal
strings and the exact unit above. Missing observations can be omitted or
set to null. Distinguish derived oil densities from measured densities in
oil_density_reference. In standard_gas_property_reference, tie the
density or the molar mass and Z to the same gas cut and common standard gas
state used for the reported volumes. Use the declared standard-state Z for
that gas cut. Gas gravity alone provides no independent standard density
unless its reference density and the report’s calculation basis are known.
Residuals and source checks
Per cubic metre of final stock-tank oil, the oil mass after stage $i$ is $B_i\rho_i$. The gas mass is $q_i\rho_{g,i}$, where $q_i$ is the independently reported gas volume on the common standard basis. The gas density can be reported directly or calculated as $M_ip_s/(Z_iRT_s)$. When both routes exist, the audit checks their intersection and reports disagreements.
The stage mass and GOR residuals are $B_{i-1}\rho_{i-1}-B_i\rho_i-q_i\rho_{g,i}$ and $R_{i-1}-R_i-q_i$. Here $R_i$ represents the gas volume still to be removed along the reported train. It is zero at the endpoint by definition even though the final oil can contain dissolved light components. The final oil volume ratio is one by definition. When a reported endpoint interval contains its exact value, the audit uses that value. A missing or incompatible endpoint does not supply a value for the corresponding balance. The raw observations and their input SHA-256 are retained.
Whole-train residuals compare feed and final oil directly and sum the reported gas withdrawals. Intermediate oil inventories cancel, so a whole train may close with missing or disagreeing intermediate values. Review each transition and the endpoint alongside the whole-train result. If oil mass loss and gas volume are strictly positive, the audit reports their ratio as an implied gas density for diagnosis. It is never fed back into a closure check. A strictly increasing oil mass is incompatible with gas removal even when the gas density is missing.
A compatible row encloses zero using outward-rounded interval propagation. Input correlations can make the true feasible region smaller. Use actual measurement uncertainties when available. Values derived from each other can provide a bookkeeping check, with independent measurements required for a physical validation claim.
Reproduce the synthetic example
The synthetic input contains a feed, an intermediate separator and a final stock tank. The documented repository test reads this exact file, verifies the input identity and checks both balances for every transition and the complete train.
import json
from pathlib import Path
from pvt_separator_mass import audit_separator_mass
path = Path("devtools/pvt_examples/separator_mass_audit.json")
report = audit_separator_mass(json.loads(path.read_text(encoding="utf-8")))
print(report["whole_train"])
Primary source screening
ResLab 70-00286
documents a two-stage separator test for Volve well 15/9-F-4, MDT bottle
6103-MA. The checked copy had SHA-256
f493b6ca4e56a6afc31089871a5325b671d5763066f3401e0f1862df016bc89f.
Use printed page 29, Table 4.6.1, for the feed and serial stage data.
Pages 30–31 report the gas compositions and molar masses, and page 32
describes the final oil. The gas reference condition of 15 °C and 1.01325
bara is taken from the same report’s standard-basis definition on page 17.
The feed density for this train is calculated as identified in the executive
summary. The single-stage experiment’s measured feed density and its
different final oil belong to that other experiment. The bottle has reported
oil-based-mud contamination.
Under half-last-digit printing bounds, both stage GOR residuals and the whole-train GOR residual enclosed zero. The first-stage oil mass loss was 61.70325–63.47355 kg per m³ of final oil. The final volume ratio printed in Table 4.6.1 excluded one. The two gas cuts have molar masses but the corresponding gas Z or independently reported standard density was not found. Keep the terminal ratio and both gas-mass balances unresolved for admission pending source clarification and gas-basis evidence. The data-use record for the specific standalone archive object remains to be verified.