Coriolis Pressure Correction Example (CMF300H)

This notebook demonstrates pressure correction of Coriolis mass flow for a real-life calibration case at an accredited calibration facility.

Case

  • Meter: Emerson Micro Motion CMF300H

  • Calibration pressure: 60 bara

  • Meter Factory calibration pressure (P_cal): 2.55 bara

  • Reference accumulated mass: 4120.487 kg

  • Coriolis accumulated mass (pressure compensation disabled): 4105.007 kg

  • Flow rate during run: 82 tonnes/hr

The CMF300H process-pressure effect from PDS is -0.006 % of rate per bar. The equation input PCm is the correction factor, i.e. opposite sign of PDS effect:

\[PCm = -\,(\text{PDS process-pressure effect}) = +0.006\;\%/\text{bar}\]

Equations

Mass-flow pressure correction:

\[m_{corr} = m_{meas}\left(1 + \frac{PCm}{100}(P_{act}-P_{cal})\right)\]

Error relative to reference meter:

\[\text{error}~[\%] = 100\cdot\frac{m - m_{ref}}{m_{ref}}\]
import pandas as pd
from pvtlib.metering import coriolis_flowmeters
# Input data
p_act_bara = 60.0
p_cal_bara = 2.55
reference_mass_kg = 4120.487
measured_mass_kg = 4105.007
flow_rate_tonnes_per_hr = 82.0

# CMF300H process-pressure effect from PDS [%/bar]
pds_process_pressure_effect_pct_per_bar = -0.006

# Equation input PCm is opposite sign of PDS process-pressure effect
pc_m_pct_per_bar = -pds_process_pressure_effect_pct_per_bar
corrected_mass_kg = coriolis_flowmeters.coriolis_massflow_corr_pres(
    m_meas=measured_mass_kg,
    P_act=p_act_bara,
    PCm=pc_m_pct_per_bar,
    P_cal=p_cal_bara,
)

error_without_correction_pct = 100.0 * (measured_mass_kg - reference_mass_kg) / reference_mass_kg
error_with_correction_pct = 100.0 * (corrected_mass_kg - reference_mass_kg) / reference_mass_kg
results = pd.DataFrame(
    [
        {
            "Condition": "Without pressure correction",
            "Accumulated mass [kg]": measured_mass_kg,
            "Error [%]": error_without_correction_pct,
        },
        {
            "Condition": "With pressure correction",
            "Accumulated mass [kg]": corrected_mass_kg,
            "Error [%]": error_with_correction_pct,
        },
    ]
)

results.round({"Accumulated mass [kg]": 3, "Error [%]": 2})
Condition Accumulated mass [kg] Error [%]
0 Without pressure correction 4105.007 -0.38
1 With pressure correction 4119.157 -0.03

This concludes that the original deviation is explained by the process pressure effect. This is the expected behavior of a Coriolis flowmeter operating at a different pressure than the pressure at which it was originally calibrated.