Coriolis Pressure Correction Example (CMF300H)¶
This notebook demonstrates pressure correction of Coriolis mass flow for a real-life calibration case at an accredited calibration facility.
Case¶
Meter: Emerson Micro Motion CMF300H
Calibration pressure: 60 bara
Meter Factory calibration pressure (
P_cal): 2.55 baraReference accumulated mass: 4120.487 kg
Coriolis accumulated mass (pressure compensation disabled): 4105.007 kg
Flow rate during run: 82 tonnes/hr
The CMF300H process-pressure effect from PDS is -0.006 % of rate per bar.
The equation input PCm is the correction factor, i.e. opposite sign of PDS effect:
Equations¶
Mass-flow pressure correction:
Error relative to reference meter:
import pandas as pd
from pvtlib.metering import coriolis_flowmeters
# Input data
p_act_bara = 60.0
p_cal_bara = 2.55
reference_mass_kg = 4120.487
measured_mass_kg = 4105.007
flow_rate_tonnes_per_hr = 82.0
# CMF300H process-pressure effect from PDS [%/bar]
pds_process_pressure_effect_pct_per_bar = -0.006
# Equation input PCm is opposite sign of PDS process-pressure effect
pc_m_pct_per_bar = -pds_process_pressure_effect_pct_per_bar
corrected_mass_kg = coriolis_flowmeters.coriolis_massflow_corr_pres(
m_meas=measured_mass_kg,
P_act=p_act_bara,
PCm=pc_m_pct_per_bar,
P_cal=p_cal_bara,
)
error_without_correction_pct = 100.0 * (measured_mass_kg - reference_mass_kg) / reference_mass_kg
error_with_correction_pct = 100.0 * (corrected_mass_kg - reference_mass_kg) / reference_mass_kg
results = pd.DataFrame(
[
{
"Condition": "Without pressure correction",
"Accumulated mass [kg]": measured_mass_kg,
"Error [%]": error_without_correction_pct,
},
{
"Condition": "With pressure correction",
"Accumulated mass [kg]": corrected_mass_kg,
"Error [%]": error_with_correction_pct,
},
]
)
results.round({"Accumulated mass [kg]": 3, "Error [%]": 2})
| Condition | Accumulated mass [kg] | Error [%] | |
|---|---|---|---|
| 0 | Without pressure correction | 4105.007 | -0.38 |
| 1 | With pressure correction | 4119.157 | -0.03 |
This concludes that the original deviation is explained by the process pressure effect. This is the expected behavior of a Coriolis flowmeter operating at a different pressure than the pressure at which it was originally calibrated.