Ultrasonic Meter Wet Gas Over-Reading Correction (Van Putten)¶
An ultrasonic meter in wet gas reads too high, because the liquid occupies part of the pipe cross section. This example demonstrates the Van Putten over-reading correction, which converts the indicated gas volume flow into the actual gas volume flow.
The notebook plots the over-reading surface, then corrects a single case using both critical gas Froude number variants: from the water-liquid ratio, Equation (27), and from the gas Ohnesorge number, Equations (25)-(26).
All input values in the notebook are illustrative.
Required packages: pvtlib, numpy, pandas, matplotlib
Format: Jupyter Notebook
Reference¶
van Putten et al., Ultrasonic Meters in Wet Gas Application, North Sea Flow Measurement Workshop, 2015. Paper
The implemented model is the simplified one, Equations (29)-(30). The full physical model, Equation (28), is not implemented.
within_JIP_envelope only confirms that the gas volume fraction,
Lockhart-Martinelli parameter, density ratio and gas Froude number are inside the
ranges reported from the JIP tests. Meter type, pipe size, installation and fluids
still require engineering assessment.